| DISPLAY | Screen | TFT Colour Screen - 7 Inch |
| Parameters | CO2,Waveform, Pleth Waveform, 5%SpCO2 Pulse Rate, EtCO2, Respiration Rate Systolic (mmHg) and Diastolic (mmHg) Pressure | |
| Sweep | 25 (mm/Sec) | |
| Bandwidth | 0.1 to 35 Hz | |
| Menu Controls | General-Freez, CO2-Alaram Settings_menu driven, SpO2-Trace on / off, HR Source select | |
| Indications | Alarm , Finger off,Zeroing Probe disconnected | |
| EtCo2 | Transducer Type | Mainstream CO2 Sensor |
| Principel of Operation | Non-dispersive infrared (NDIR) | |
| CO2 Range | 0-150 mmHg | |
| CO2 Resolution | 0-49mmHg ±0.1 mmHg 50-150mmHg ±0.2 mmHg | |
| CO2 Accuracy | ±2mmHg @ <5.0% CO2 (at BTPS) <10% of reading @>5.0% CO2, (at BTPS) | |
| Response Time | 200 msecs | |
| %SpO2 | Read Out | 0 to100% |
| Accuracy | ±2% at 70 to 100% SpO2,±3% at 50 to 69% SpO2 | |
| Plethysmograph | Displayed with 14Bit Ladder | |
| Wave Gain | Auto Gain | |
| Alarm Limit Adjust | Low -> 50-100% SpO2,High -> 80-10% SpO2 | |
| Alarm Limit Resolution | ±1% SpO2 | |
| NIPB | Measurement Method | Oscillometric Method |
| Adult | 20 to 300mmHg | |
| Accuracy | ±5mmHg | |
| Cuff Inflation Time | 30 to 20 Seconds | |
| Cuff Deflation Time | 20 to 50 Seconds | |
| Over Pressure | 320mmHg ±15mmHg | |
| PULSE | Read Out | 30 to 250 BPM |
| Accuracy | ±2% + 1 BPM | |
| Average Technique | Moving Average Window | |
| Alarm Limit Adjust | Low -> 0 -150 BPM -High -> 40 -250 BPM | |
| Alarm Limit Resolution | ±1BPM | |
| INDICATIONS | Visual | Sensor OFF, Finger Off, Check Probe, Searching for Pulse, Bp Measurement Error, BP Over Pressure Low & High, Alarm Violation and Auto Power Save Indication |
| Audio | Variable Pitch Beep for every pulse. Continuous Dual Beep for Alarm Violation, Intermittent beep for Sensor Off & Finger OFF | |
| RESPIRATION | Rate Range | 0 to 150 Breaths per Minute (BPM) |
| Accuracy | ±1 breath | |
| CALIBRATION | No routine user calibration required. Airway Adapter Zero is required when changing to a different style of airway adapter. | |
| ALARMS | Alarm Limit Settings | Independent of all Parameters Last set values are stored in Non-volatile Memory |
| Alarm Accuracy | ±1 | |
| Audio / Visual | Low & High Alarm Violation Continuous Beep for Alarm Violation | |
| BATTERY | Type | Li-ion,14.6 V,2.0 Ah |
| Charge Time | 16 Hours | |
| Operation Time | 2Hrs | |
| GENERAL | Dimension | 150mm (H) × 240mm (L) × 120mm(w) |
| Weight | 2 Kgs | |
| Power Input | 230V ±10% AC, 50Hz | |
| Consumption | 25VA | |
| Weight | 2 Kgs | |
| Earth Leakage Current | <100 micro A | |
| Electrical Safety | Meets IEC 6060 1-1 Standards | |
| Operating Temperature | +5C to +45C at 95% RH-NON Non Condensing |